(1) MWS is, WFI production volume also proportionally increases in proportion to the pressure of the primary steam. Typically, in water sampling operation are sent 5 to 15% more water than the supplied WFI production volumes. Supply water, I have to limit the flow rate an orifice from evaporator bottom so that it does not flow in excess to the next can in a pipe entering the next evaporator. By properly selecting the diameter of the orifice, when the feed water enters the excess MWS, detects the first high level in the first evaporator to issue an alarm. Even if the first evaporator has a high level, the second evaporation later cans and by adjusting the orifice diameter of the saturated water outlet of each can so as not to high water. Each evaporator is to be able to visually confirm the presence or absence of saturated water provided a viewing window on the bottom.